Teledyne analytical instruments, 4 sampling – Teledyne 7500 - Infrared gas analyzer - operating manual User Manual

Page 12

Advertising
background image

12

Model 7500 Instruction Manual

Teledyne Analytical Instruments

Observe the following when connecting the gas pipes.

• The pipes should be connected to the gas inlet and outlet at the rear panel of the analyzer, respectively.

• Connect the sampling system to the instrument by using corrosion-resistant tube such as Teflon,

stainless steel, or polyethylene. In case where there is no danger of corrosion, donʼt use rubber or soft

vinyl tube. Analyzer indication may become inaccurate due to the adsorption of gases.

• Piping connections are Rc1/4 (NPT1/4) female-threaded. Cut the pipe as short as possible for quick

response. Pipe of ø 4mm (inside diameter) is recommendable.

• Entry of dust in the instrument may cause operation fault. Use clean pipes and couplings.

Sampling gas inlet:

Connect the pipe so that zero/span calibration standard gas or measured gas

pre-treated with dehumidification is supplied properly. The gas flow rate should

be kept constant within the range of 1 L/min ±0.5 L/min.

Sampling gas outlet: Measured gas is exhausted after measurement.

Connect the pipe so that the gas may escape through the gas outlet into the

atmosphere.

Purge gas inlet:

It is used for purging the inside of the total gas analyzer. When the analyzer must

be purged, refer to Item 3.4, Purging inside Analyzer. Use dry gas N2 or

instrumentation air for purge gas. (flow rate of 1 L/min or more should be used

and no dust or mist is contained).

3.4 Sampling

(1) Conditions of sample gas

1.

The dust contained in sample gas should be eliminated completely with filters. The filter at the

final stage should be capable of eliminating dust of 0.3 microns.

2.

The dew point of sample gas must be lower than the ambient temperature for preventing

formation of drain in the analyzer. If water vapor is contained in sample gas, its dew point should be

reduced down to about 0°C through a dehumidifier.

3

If SO3 mist is contained in sample gas, the mist should be eliminated with a mist filter, cooler, etc.

Eliminate other mist in the same way.

4.

If a large amount of highly corrosive gas such as Cl2, F2 or HCl is contained in sample gas, the

service life of analyzer will be shortened. So, avoid such gases.

5.

Sample gas temperature is allowed within a range from 0 to 50°C. Pay attention not to flow hot

gas directly into the analyzer.

Advertising