Model 7500 Instruction Manual

Teledyne Analytical Instruments

9. SPECIFICATIONS

9.1 Specifications

(1) Standard specifications

Principle of measurement:

CO2, CO, CH4, SO2; Non-dispersion infrared-ray

absorption method; Single light source and single beam

(single beam system)

O2: Paramagnetic method (O2 sensor built in) or

zirconia sensor method (O2 sensor externally installed)

Measurable gas components and measuring range:

Minimum range Maximum

range

CO2

0 to 500 ppm

0 to 100 vol%

CO

0 to 200 ppm

0 to 100 vol%

CH4

0 to 1000 ppm

0 to 100 vol%

SO2

0 to 1000 ppm

0 to 5000 ppm

O2 (built in)

0 to 5 vol%

0 to 100 vol%

O2 (external

zirconia)

0 to 5 vol%

0 to 25 vol%

• Max. 4 components measurement including O2.

• 1 or 2 measuring range per component.

• Measuring range ratio ≤ 1:5 (except built in O2) ≤ 1:20

(built in O2)

Max. 4 components and 2 ranges are selectable including

an O2 measurement. For measurable components and

possible combinations of measuring ranges, refer to

Tables 1 to 6.

Measured value indication:

Digital indication in 4 digits (LCD with back light)

• Instantaneous value of each component

• Instantaneous value after O2 correction (only in CO,

SO2 measurement with O2)

• Average value after O2 correction (only in CO, SO2

measurement with O2)

• O2 average value

Analog output signals:

4 to 20mA DC or 0 to 1V DC, non-isolated output. Analog

output corresponds to measured value indication in 1:1.

Permissible load; 550Ω max. for 4 to 20 mA DC100kΩ

min. for 0 to 1V DC

* Refer to Table 6, for the channel No. of displayed values

and analog output signals.

Analog input signal:

For signal input from externally installed O2 sensor.

Signal requirement; (1) Signal from Teledyneʼs Zirconia

O2 sensor (TYPE: ZFK7) (2) 0 to 1V DC from an O2

sensor Input section is not isolated. This feature is

effective when an O2 sensor is not built in.

Relay contact output:

1a contact (250V AC/2A, resistive load) Instrument error,

calibration error, range discrimination, auto calibration

status, solenoid valve drive for auto calibration, pump

ON/OFF.

1c contact (250V AC/2A, resistive load) Upper/lower

alarm contact output. Peak count alarm contact output.

* All relay contacts are isolated mutually and from the

internal circuit.

Contact input:

Non-voltage contact (ON/0V, OFF/5V DC, 5mA flowing at

ON); Remote range changeover, auto calibration remote

start, remote holding, average value resetting; Isolated

from the internal circuit with a photocoupler. Contact

inputs are not isolated from one another.

*Only M3.5 screw terminals are used for all signal inputs

and outputs.

Power supply:

Rated voltage:

100V to 240V AC

Operating voltage:

85V to 264V AC

Rated frequency:

50/60 Hz

Power consumption: 70 VA max.

Power inlet: Conformity to EN60320 Class I type

Operation conditions:

Ambient temperature: -5°C to 45°C

Ambient humidity:

90% RH max.

Storage conditions:

Ambient temperature: -20°C to 60°C

Ambient humidity:

100% RH max.noncondensing

Dimensions (H x W x D):

19-inch rack mounting type:

177 x 483 x 493mm

Desk-top type:

194 x 483 x 493mm

Mass: Approx. 10 kg

Finish color:

Front panel; Off-white (Munsell 10Y7.5 / 0.5 or equivalent)

Casing: Steel-blue

Enclosure: Steel casing, for indoor use

Material of gas-contacting parts:

Gas inlet/outlet; SUS304

Sample cell; SUS304/neoprene rubber

Infrared-ray transmitting window; CaF2

Internal tubing; Toaron tube

Gas inlet/outlet:

Rc1/4 or NPT1/4 internal thread

Purge gas flow rate:

1 L / min (when required)