Model 7500 Instruction Manual

Teledyne Analytical Instruments

(4) Contact input (DI): terminal block 2 (13 to 20), terminal block 3 (5 to 10)

• It is for a contact input at no voltage. An input is provided when switching to short circuit (on) or

open (off).

• No voltage is applied to the terminals

(5) Contact output (DO): terminal block 3 (13 to 20), terminal block 4 and terminal block 5

• Contact rating: 250VAC / 2A, load resistance

• An output is for a relay contact output. An output is provided when switching to conductive (on)

or open (off).

Note: The wires of analog output signals, O2 sensor input and contact input should be fixed separately

from power supply wiring and contact output wiring.

(6) Communication interface: connector 2