Model 7500 Instruction Manual

Teledyne Analytical Instruments

6.2 Calibration setting

This mode is used to set calibration concentration and

actions. The calibration setting involves calibration

concentration, zero calibration, calibration range and

auto-calibration component.

6.2.1 Setting of calibration concentration

It allows you to set concentrations of the standard gas

(zero and span) of each channel used for calibration.

1. During measurement, press the MODE key to display

the User mode.

2. Point the cursor to “Setting about Calibration” by

pressing the UP or DOWN key. Press the ENT key.

3. In the “Setting about Calibration” screen that appears,

point the cursor to “Calibration Value” by pressing the UP

or DOWN key. Press the ENT key.

4. In the “Calibration Concentration CH Selection”

screen that appears, point the cursor to CH you want to

set by using the UP or DOWN key. Press the ENT key.