Teledyne 7500 - Infrared gas analyzer - operating manual User Manual

Page 51

Advertising
background image

51

Model 7500 Instruction Manual

Teledyne Analytical Instruments

6.6 Peak alarm setting

When the number of peaks of which CO concentration

exceeds the upper limit value exceeds the setting time, a

peak alarm is outputted.

This section describes how to perform various settings of

peak alarm.

Note: The setting is optional and is valid only when peak

alarm function is provided.

1. Press the MODE key in the Measurement mode, and

the User mode appears.

2. Point the cursor to “Setting of Peak Alarm” by pressing

the or key. Press the ENT key.

3. In the “Peak Alarm Setting” item selection screen

that appears, point the cursor to any item you want by

pressing the UP or DOWN key. Press the ENT key.

4. Then, enter numeric values and perform the setting.

Entering or setting the numeric values should be carried

out by using the UP or DOWN key.

After setting, press the ENT key, and the set values you

entered are saved.

Description of setting items

• Peak Alarm: Setting of peak alarm is performed with

ON/OFF.

• Alarm value: If peak concentrations exceed the set

alarm value, a peak counter counts 1 time.

• Alarm Count: When a peak in excess of the setting time

occurs, a peak count alarm output is provided.

• Hysteresis: To prevent possible chattering when the

peak value may exceed the set peak concentration by

only 1 time, the peak count has an allowance in the

hysteresis width.