8. TROUBLESHOOTING

8.1 Error message

If errors occur, the following contents are displayed.

Error display

Error contents

Probable causes

Error No. 1

Sector motor rotation detector signal is faulty.

• Infrared ray light source is faulty.

• Sector motor rotation is faulty or stopped.

• Motor rotation detector circuit is faulty.

• Amplifier circuit is faulty.

Error No. 3

A/D conversion signal is faulty.

• Circuit is faulty.

Error No. 4

Zero calibration is not within the allowable

range.

• Zero gas is not supplied.

• Zero point is deflected much due to a dirty cell.

• Detector is faulty.

Error No. 5

A amount of zero calibration (indication value) is

over 50% of full scale.

Error No. 6

Span calibration is not within the allowable

range.

• Span gas is not supplied.

• Calibrated concentration setting does not

match cylinder concentration.

• Zero calibration is not performed properly.

• Span is deflected much due to dirty cell.

• Detector sensitivity is deteriorated.

Error No. 7

A amount of span calibration (difference

between indication value and calibrated

concentration value) is over 50% of full scale.

Error No. 8

Measured values fluctuate to much during zero

and span calibration.

• Calibration gas is not supplied.

• Time for supplying calibration gas is not short.

Error No. 9

Calibration is abnormal during auto calibration.

• Error corresponding to No. 4 to No. 8 occurred

during auto calibration.

Error No. 10

Output cable connection is improper.

• Wiring is detached between analyzer and

interface module.

• Wiring is disconnected between analyzer and

interface module.

Note: When errors No. 1, No. 3 and No. 10 occur, instrument error output contacts are conductive.

When errors No. 4 to No. 9 occur, calibration error output contacts are conductive.

Screen display and operation at the occurrence of error.

In case of Error No. 1 to No. 4, No. 6, No. 8 to No. 10