Teledyne analytical instruments – Teledyne 7500 - Infrared gas analyzer - operating manual User Manual

Page 37

Advertising
background image

37

Model 7500 Instruction Manual

Teledyne Analytical Instruments

5. In the “Calibration Selection” screen that appears, select “Both” or “Current” by pressing the UP or

DOWN key.

• When selecting “both”, Range 1 and Range 2 of the set CH

are calibrated together.

• When selecting “Current”, the range alone displayed at the

set CH is calibrated.

To close “Setting of Calibration Range”

To close “Setting of Calibration Range” or to cancel this mode

midway, press the ESC key. A previous screen will return.

EXAMPLE:

CH1

CO2

Range 1: 0 to 10 vol%

Range 2: 0 to 20 vol%

both

CH2

CO

Range 1: 0 to 500 ppm

Range 2: 0 to 2000 ppm

current

CH1: Range 1 and Range 2 are calibrated together, with zero or span calibration.

CH2: Only currently displayed range is calibrated, with zero or span calibration.

Manual Calibration screen

In case of “both” setting

Two cursors will appear in both ranges.

Note: When calibration is performed by the “Both” setting under the normal operating condition, prepare a

span gas cylinder on the normal operating range side. It is recommend to perform span gas calibration in

the normal operating range. The other range that is calibrated by “Both” may result in some error (max. of

±5% of FS).