6.4.2 Forced stop of auto calibration

This mode is used to cancel the auto calibration forcedly.

1. In the Menu mode that is displayed, point the cursor

to “Setting of Auto Calibration” by pressing the UP or

DOWN key. Press the ENT key.

2. In the “Setting of Auto Calibration” item selection

screen that appears, point the cursor to “Setting of Auto

Calibration” by pressing the UP or DOWN key. Press the

ENT key.

3. “Stop Auto Calibration” is inverted. A message

appears, prompting you to verify that you want to

cancel or continue auto calibration. To cancel the auto

calibration, press the ENT key. If you press the ESC key,

auto calibration is not stopped.