5.3 Overview of display screen

(1) Measurement mode screen

On turning on the power switch, the Measurement Mode screen will appear.

The measurement screen depends on the number of components. The following screen configuration

shown as an example is for CO2, CO, O2 (Output at channel 6).

When channel 5 or later is displayed, scroll up or down key to view another configuration which is beyond

the screen.

No. Name

Function

No. Name

Description

1

Component

display

Displays component

of instantaneous

value, converted

instantaneous value,

converted average

value, etc.

5

Peak alarm

component

display

Displays peak alarm

component.

2

Concentration

display

Displays measured

value of concentration.

6

Peak alarm

concentration

display

Displays peak alarm

concentration (Upper

limit value).

3

Range display

Displays range

values.

7

Peak alarm count Displays the alarm

times exceeding the

peak value.

4

Unit display

Displays unit with ppm

and vol%.

8

Peak alarm unit

display

Displays unit of peak

alarm with times/h.

Fig. 5-3 Name and function of measurement mode screen

Instantaneous value and concentration value:

The concentration display of CH (component) where sampling components such as “CO2”, “CO” or

“O2 are displayed in the component display, indicates current concentration values of the measured

components contained in gas that is now under measurement.

• O2 correction concentration values:

CH (components) where “cv**” is displayed as “cv CO” in the component display are calculated from the

following equation, by setting sampling components, O2 instantaneous / concentration values and O2

correction reference value (see item 6.8).

Conversion output=

On: The value of the O2 correction reference value (Value set by application)

Os: Oxygen concentration (%)

Cs: Concentration of relevant measured component

K: The value of the fractional part is this equation.

Where, K is When K ≥ 4, K=4. When K < 0, K=4. When Cs < 0, K=0.

The correction components are CO and SO2 only.