Model 7500 Instruction Manual

Teledyne Analytical Instruments

(2) Sampling gas flow rate

A flow rate of sampling gas must be 1 L/min ±0.5 L/min. A flow meter should be provided as

shown in Fig. 3-2 to measure flow rate values.

(3) Preparation for standard gas

Prepare the standard gas for zero/span calibration.

No O2 meter

Built in O2 meter

Externally mounted

zirconia O2 meter

Zero gas

N2 gas

N2 gas

Dry air or air (No

calibration performed

with CO2 meter)

Span gas except for O2 Gas with concentration

of 90% or more of full

scale

Gas with concentration

of 90% or more of full

scale

Gas with concentration

of 90% or more of full

scale

Span gas for O2

Gas with concentration

of 90% or more of full

scale

1 to 2% of O2 gas

(4) Purging of instrument inside

The inside of instrument need not be purged generally except for the following cases.

1.

A combustible gas component is contained in sample gas.

2.

Corrosive gas is contained in the atmospheric air at the installation site.

3.

The same gas as the sample gas component is contained in the atmospheric air at the installation

site.

In such cases as above, the inside of analyzer should be purged with the air for instrumentation or N2.

Purging flow rate should be about 1L/min.

If dust or mist is contained in purging gas, it should be eliminated completely in advance.

(5) Pressure at sampling gas outlet

Pressure at the sampling gas outlet should be set to atmospheric pressure.