6.2.3 Setting of calibration range

This mode is used to set if the range of each CH

(component) at the calibration manual calibration or auto

calibration) should be calibrated with a single range or 2

ranges.

1. During measurement, press the MODE key to display

the User mode.

2. Point the cursor to “Setting about Calibration” by

pressing the UP or DOWN key. Press the ENT key.

3. In the “Setting about Calibration” screen that appears,

point the cursor to “About Calibration Range” by pressing

the UP or DOWN key. Press the ENT key.

4. In the “Calibration Range CH Selection” screen that

appears, point the cursor to the CH you want to set by

pressing the UP or DOWN key. Press the ENT key.