Teledyne 7500 - Infrared gas analyzer - operating manual User Manual

Page 47

Advertising
background image

47

Model 7500 Instruction Manual

Teledyne Analytical Instruments

6.5 Setting of auto zero calibration

6.5.1 Auto zero calibration

Auto zero calibration is automatically carried out when

zero calibration are set.

Components for which a calibration is to be made are

determined by setting of auto calibration component in

6.2.4.

Before changing the setting of auto zero calibration, set

the ON/OFF to OFF.

1. During measurement, press the MODE key to display

the User mode.

2. Point the cursor to “Setting of Auto Zero Calibration”

by pressing the UP or DOWN key. Press the ENT key.

3. In the “Setting of Auto Zero Calibration” screen that

appears, point the cursor to any item you want by

pressing the UP or DOWN key. Press the ENT key.

4. In the “Auto Zero Calibration Parameter Setting”

screen that appears, perform the value entry or the

setting. For the value entry or setting change, use the

UP or DOWN key. To change the setting, use the RIGHT

key to move the cursor to the right.

After setting, press the ENT key, and auto zero

calibration is carried out by the entered setting value.

Description of setting items

• Start Time: Setting at the first calibration (day of the

week, hour, minute)

• Cycle: A period between the start time of one

calibration and another (unit : hour/day)

• Flow Time: The time required for the calibration gas to

be replaced in the cell

• ON/OFF: Auto zero calibration ON or OFF

To close “Auto Zero Calibration”

To close the “ Auto Zero Calibration “ or cancel this

mode midway, press the ESC key. A previous screen will

return.