Model 7500 Instruction Manual

Teledyne Analytical Instruments

6.3.2 Hysteresis setting

To prevent chattering of an alarm output near the alarm

setting values, set hysteresis.

1. In the “Alarm Setting CH Selection” screen that

appears, point the cursor to “Hysteresis” by pressing the

UP or DOWN key. Press the ENT key.

2. In the “Hysteresis Value Setting” screen that appears,

enter hysteresis values. For the value entry, 1-digit value

is increased or decreased by pressing the UP or DOWN

key, and pressing the RIGHT key moves the digit. After

setting, press the ENT key to carry out hysteresis.

To close Hysteresis Setting

To close the “Hysteresis Setting” or cancel the mode

midway, press the ESC key. A previous screen will

return.

Setting range

0 to 20% of full scale

A full scale means each range provides a full scale of

width.

Caution

The hysteresis is common to all CHs (components).

Hysteresis

If hysteresis values exceed the upper limit value as

shown in graph, an alarm output is provided. Once the

alarm output is turned ON, it remains ON until the value

falls below the set lower limit of the hysteresis indication.