43

Model 7500 Instruction Manual

Teledyne Analytical Instruments

6.4 Setting of auto calibration

6.4.1 Auto calibration

Auto calibration is automatically carried out when zero

calibration and span calibration are set.

Before changing the setting of auto calibration, set the

ON/OFF to OFF.

1. During measurement, press the MODE key to display

the User mode.

2. Point the cursor to “Setting of Auto Calibration” by

pressing the or key. Press the ENT key.

3. In the “Setting of Auto Calibration” screen that

appears, point the cursor to any item you want by

pressing the UP or DOWN key. Press the ENT key.

4. In the “Auto Calibration Parameter Setting” screen

that appears, perform the value entry or the setting. For

the value entry or setting change, use the UP or DOWN

key. For the value entry and setting, use the UP or

DOWN key. To change the setting, use the RIGHT key to

move the cursor to the right.

After setting, press the ENT key, and auto calibration is

carried out by the entered setting value.

Description of setting items

• Start Time: Setting at the first calibration (day of the

week, hour, minute)

• Cycle: A period between the start time of one

calibration and another (unit : hour/day)

• Flow Time: The time required for the calibration gas to

be replaced in the cell

• ON/OFF: Auto zero calibration ON or OFF

To close Setting of Auto calibration

To close the “Setting of Auto calibration” or cancel this

mode midway, press the ESC key. A previous screen will

return.