Teledyne analytical instruments – Teledyne 7500 - Infrared gas analyzer - operating manual User Manual

Page 60

Advertising
background image

60

Model 7500 Instruction Manual

Teledyne Analytical Instruments

6.9.2 Span calibration

It is used to perform a span point adjustment. Supply

calibration gas with concentration set to the span value

to perform the span calibration. For the span calibration

gas for the SO2, CO2, CO, H4 measurement, use the

standard gas with a concentration of 90% or more of

the range value. For the calibration gas for the O2

measurement, use the standard gas of about 2 vol%.

1. Press the SPAN key on the Measurement screen to

display the Manual Span Calibration screen.

2. Press the UP or DOWN key to select CH (component)

to be calibrated. After selection, press the ENT key, and

the span calibration contact output on corresponding to

CH is turned ON. And supply span gas.

CAUTION

When Range Interlock from “Calibration Range” of

the Calibration Setting mode is set, span calibration is

performed together with 2 Ranges.

3. Wait until the indication is stabilized in the state where

the calibration gas is supplied. After the indication has

been stabilized, press the ENT key. Span calibration of

Range selected by the cursor is performed.

To close Span Calibration

To close the “ Span Calibration “ or cancel this mode

midway, press the ESC key. A previous screen will

return.