33

Model 7500 Instruction Manual

Teledyne Analytical Instruments

5. In the “Calibration Concentration Selection” screen

that appears, select any concentration item you want to

set by pressing the UP, DOWN, and RIGHT key.

Note: Analyzers other than the zirconia O2 instrument

cannot perform zero setting.

6. In the “Calibration Concentration Value Setting”

screen that appears, enter calibration gas concentration

values (zero and span). For value entry, press the UP or

DOWN key, and a 1-digit value increases or decreases.

By pressing the RIGHT key, the digit moves.

After setting, save the entry by pressing the ENT key.

The saved value becomes valid from the next calibration

process.

Note: Enter the set values corresponding to each range.

When the O2 measurement uses atmospheric air for the

zero gas, set the concentration value to 21.00. When

the cylinder air is used, set to the concentration value as

indicated on the cylinder.

To close the setting

To close the calibration concentration value setting

process or cancel this mode midway, press the ESC key.

A previous screen will return.

Setting range of values

Paramagnetic O2, CO2, CO, SO2 and CH4

measurement:

Zero gas:

Fixed at 0

Span gas:

Minimum digit, 1 to 100% of full

scale

(Full scale (FS) is the same as each range

value.)

Zirconia O2 measurement :

Zero gas:

5 to 25 vol%

Span gas:

0.01 to 5 vol%

The setting cannot be performed beyond the range.