Terminal block 4 <TN4>

Between 1 – 2: CH4 span calibration contact output

Between 3 – 4: CH3 span calibration contact output

Between 5 – 6: CH2 span calibration contact output

Between 7 – 8: CH1 span calibration contact

Between 9 – 10: Zero calibration contact output When the

calibration contact output is measured with

manual calibration, the calibration contact

corresponding to calibration channel is

conductive.

For the automatic calibration, they are worked sequentially

according to “6.4 Auto calibration setting”. If calibration is not

performed, all of them are open.

Between 11 – 12: For internal connection. Must not be wired.

(Must not be used as junction terminal.)

Between 13 – 14: Automatic calibration in progress,

contact output.

Conductive during automatic calibration. Open

otherwise.

Between 15 – 16: Calibration error contact output. Conductive

when error is produced at zero or span

calibration. Normally open.

Between 17 – 18: Pump ON/OFF contact output. (Used for

turning ON/OFF the pump. Conductive during

measurement and open at zero span

calibration.)