Teledyne analytical instruments 6.9 calibration – Teledyne 7500 - Infrared gas analyzer - operating manual User Manual

Page 59

Advertising
background image

59

Model 7500 Instruction Manual

Teledyne Analytical Instruments

6.9 Calibration

6.9.1 Zero calibration

It is used for zero point adjustment. For zero calibration

gas, see 3.3 (3), Preparation for standard gas in

Sampling. Use a gas according to application.

1. Press the ZERO key on the Measurement screen to

display the Manual Zero Calibration screen.

2. Select the CH (component) to be calibrated by

pressing the UP or DOWN key. After selection, press the

ENT key, and zero gas will be supplied.

CAUTION

For the CH (components) that is set to “at once” in the

“Zero Calibration” of the Calibration Setting mode, zero

calibration is also carried out at the same time.

3. Wait until the indication is stabilized with the zero gas

supplied. After the indication has been stabilized, press

the ENT key. Zero calibration in Range selected by the

cursor is carried out.

To close Zero Calibration

To close the “Zero Calibration “ or cancel this mode

midway, press the ESC key. A previous screen will

return.