(2) Standard functions

Output signal holding:

Output signals are held during manual and auto

calibrations by activation of holding (turning on its

setting). The values to be held are the ones just before

start calibration mode. Indication values will not be held.

Remote output holding:

Output signal is held at the latest value by short-circuiting

the remote output holding input terminals. Holding

is maintained while the terminals are short-circuited.

Indication values will not be held.

Remote range changeover:

Measuring range can be changed according to an

external signal when remote range changeover input

is received. Changeover is effective only when remote

range setting is turned on. In this case, measuring range

cannot be changed manually.

When the contact input terminals for each component

are short-circuited, the first range is selected, and it is

changed over to the second range when the terminals are

open.

Range identification signal:

The present measuring range is identified by a contact

signal. The contact output terminals for each component

are short-circuited when the first range is selected, and

when the second range is selected, the terminals are

open.

Auto calibration:

Auto calibration is carried out periodically at the preset

cycle. When a standard gas cylinder for calibration and

a solenoid valve for opening/closing the gas flow line are

prepared externally by the customer, calibration will be

carried out with the solenoid valve drive contacts for

zero calibration and each span calibration turned on/off

sequentially at the set auto calibration timing.

Auto calibration cycle setting:

Auto calibration cycle is set. Setting is variable within 1

to 99 hours (in increments of 1 hour) or 1 to 40 days (in

increments of 1 day).

Gas flow time setting:

The time for flowing each calibration gas in auto

calibration is set. Settable within 60 to 599 seconds (in

increments of 1 second)

Auto calibration remote start:

Auto calibration is carried out only once according to an

external input signal. Calibration sequence is settable in

the same way as the cyclic auto calibration. Calibration

starts when a non-voltage rectangular wave is applied to

the auto calibration remote start input terminals (opened

after short-circuiting for 1.5 seconds or longer). Auto

calibration is started when the contacts open.

Auto zero calibration:

Auto zero calibration is carried out periodically at

the preset cycle. This cycle is independent on “Auto

calibration” cycle.

When zero calibration gas and solenoid valve for

opening/closing the calibration gas flow line are prepared

externally by the customer, zero calibration will be carried

out with the solenoid valve drive contact for zero

calibration turned on/off at the set auto zero calibration

timing.

Auto zero calibration cycle setting:

Auto zero calibration cycle is set. Setting is variable

within 1 to 99 hours (in increments of 1 hour) or setting is

variable within 1 to 40 days (in increments of 1 day).

Gas flow time setting:

The timing for flowing zero gas in auto zero calibration

is set. Settable 60 to 599 seconds (in increments of 1

second)

Upper/lower limit alarm:

Alarm contact output turns on when the preset upper or

lower limit alarm value is reached.

Contacts close when the instantaneous value of each

component becomes larger than the upper alarm limit

value or smaller than the lower alarm limit value.

Instrument error contact output:

Contacts close at occurrence of analyzer error No. 1, 3 or

10.

Calibration error contact output:

Contacts close at occurrence of manual or auto

calibration error (any of errors No. 4 to 9).

Auto calibration status contact outputs:

Contacts close during auto calibration.

Pump ON/OFF contact output:

During measurement, this contact close. While calibration

gas is flowing, this contact open. This contact is

connected in power supply of pump, and stop the sample

gas while calibration gas flowing.

(3) Optional functions

O2 correction:

Conversion of measured CO and SO2 gas concentrations

into values at standard O2 concentration

21-On

Correction formula: C = ––––––––- x Cs

21-Os

C: Sample gas concentration after O2 correction