Average period

It allows you to set an average period of the average value of

O2 correction.

It enables you to set an average time of 1 to 59 minutes (1-

minute step) or 1 to 4 hours (1-hour step).

Changing the setting resets the average value of O2

correction and O2 average value. (Pressing the ENT

validates the resetting only for components whose setting

was changed.)

Example of average action

Suppose the average period is 1 hour.

• Sampling occurs every 30 seconds

• Every 30 seconds, the average for last 1 hour (time setting) is output.

• At the instant of resetting, zero is assumed for all past values. It means the average value will not be

correct for 1 hour after resetting.

Maintenance mode

To open the maintenance mode, enter a password. After entering the password, press the ENT key.

The password can be used for the Password Setting in the Maintenance mode. A password is set to

“0000” before factory-shipment. This value is available for the Maintenance mode.