Teledyne analytical instruments – Teledyne 7500 - Infrared gas analyzer - operating manual User Manual

Page 72

Advertising
background image

72

Model 7500 Instruction Manual

Teledyne Analytical Instruments

CS: measured concentration of sample gas

Os : Measured O2 concentration

On: Standard O2 concentration (value changeable by

setting)

*The upper limit value of the fractional part in this

calculation is 4. The result of calculation is indicated and

output in an analog output signal.

Average value after O2 correction and O2 average

value calculation:

The result of O2 correction or instantaneous O2 value

can be output as an average value in the determined

period of time. Used for averaging is the moving average

method in which sampling is carried out at intervals of 30

seconds.

(Output is updated every 30 seconds. It is the average

value in the determined period of time just before the

latest updating.) Averaging time is settable within 1 to 59

minutes (in increments of 1 minute) or 1 to 4 hours (in

increments of 1 hour).

Average value resetting:

The above-mentioned output of average value is started

from the initial state by applying a non-voltage rectangular

wave to the average value resetting input terminals

(opened after short-circuiting for 1.5 seconds or longer).

Output is reset by short-circuiting and restarted by

opening.

CO concentration peak count alarm:

(added only for CO/O2 measurement) Alarm output turns

on according to the preset concentration and count.

Whenever the instantaneous value of CO exceeds the

preset concentration value, count increments. If the count

exceeds the preset value in one hour, the alarm contacts

close.

Communication function:

RS-232C (9pins D-sub)

Half-duplex bit serial

Start-stop synchronization

ModbusTM protocol

Contents: Read/Write parameters; Read measurement

concentration and instrument status.

Remark : When connecting via RS-485 interface, an RS-

232C / RS-485 converter should be used.

(4) Applicable standards

Product safety: EN61010-1 :2001

EMC: EN61326-1 :1997, A1 :1998, A2 :2001

(5) Performance

Repeatability: ±0.5% of full scale

Linearity:

±1% of full scale

Zero drift:

±2% of full scale/week

Span drift:

±2% of full scale/week

Response time: (for 90% FS response, including

replacement time of sample gas)

1 or 2 component measurement: Within 15 seconds

3 or 4 component measurement : Within 30 seconds

Interference from other gases:

Interference

component

CO2

ana-

lyzer

CO

analyzer

CH4

ana-

lyzer

SO2

ana-

lyzer

O2

ana-

lyzer

CO 1000

ppm

≤1%

FS

≤1%

FS

≤1%

FS

CO2 15%

≤1% FS (for

200 ppm

analyzer,

≤2.5% FS)

≤1%

FS

≤1%

FS

≤2%

FS

H2O satura-

tion at 20°C

≤1%

FS

≤1% FS (for

200 ppm

analyzer,

≤2.5% FS)

≤1%

FS

H2O satura-

tion at 2°C

≤2.5% FS

(for 200

ppm ana-

lyzer)

≤5%

FS

(6) Standard Requirements for Sample Gas

Flow rate:

1L / min ±0.5L / min

Temperature:

0 to 50°C

Pressure:

10 kPa or less (Gas outlet side should

be open to the atmospheric air.)

Dust: 100 µg/Nm3 in particle size of 0.3 µm or less

Mist:

Unallowable

Moisture:

Below a level where saturation occurs

at room temperature (condensation unallowable). Below

a level to cause saturation at 2°C for CO measurement

with 0 to 200 ppm range and SO2 measurement.

Corrosive component:

HCl 1 ppm or less

Standard gas for calibration:

Zero gas; Dry N2

Span gas; Each sample gas having concentration 90

to 100% of its measuring range (recommended). Gas

beyond concentration 100% is unusable.

In case a zirconia O2 analyzer is installed externally

and calibration is carried out on the same calibration gas

line:

Zero gas; Dry air or atmospheric air

Span gas; For other than O2 measurement, each

sample gas having concentration 90 to 100% of its

measuring range. For O2 measurement, O2 gas of 1 to 2

vol%.

(7) Installation conditions

Use this instrument indoors.

(Select a place where the equipment does not receive

direct sunshine, draft/rain or radiation from hot

substances. If such a place cannot be found, a roof or

cover should be prepared for protection. )

A vibration-free place.

A place which is clean around the analyzer.