Teledyne analytical instruments – Teledyne 7500 - Infrared gas analyzer - operating manual User Manual

Page 52

Advertising
background image

52

Model 7500 Instruction Manual

Teledyne Analytical Instruments

Setting range

• Alarm Value: 0 to 1000 ppm 5 ppm step (initial value: 500 ppm)

• Alarm Count: 1 to 99 times (initial value: 5 times)

• Hysteresis: 0 to 20% of full scale (initial value: 0% of full scale)

* The hysteresis setting is made in terms of full scale.

Action of peak alarm

If CO concentration exceeds the alarm value, counting will begin. If the number of peaks is over the set

times per hour, a peak count alarm becomes closed (ON). If it is less than the set times per hour, it is

open (OFF). Since 5 times of peaks /hour is marked at 1 section from the above graph, the peak count

alarm is turned ON. Since peaks of more than 5 times per 1 hour occur at the interval between 1 and 2 ,

the peak count alarm remains ON. Since peaks are reduced to 4 times per hour, it is turned OFF.

Like the hysteresis of the alarm setting , the hysteresis prevents possible chattering when measured gas

is fluctuated near the alarm value.

* For 10 minutes after the power is turned ON, a peak alarm counting is not carried out.

Releasing peak alarm

To release the peak alarm, set the peak alarm to OFF.

Turning on the peak alarm initiates counting from 0.