35

Model 7500 Instruction Manual

Teledyne Analytical Instruments

5. In the “Manual Calibration Selection” screen that appears, select “At once” or “Each” by pressing the

UP or DOWN key. When selecting “Both”, the CH (components) to be set can be zero-calibrated at the

same time. When selecting “Each”, either of the CH (components) to be selected is zero-calibrated. After

setting, press the ENT key, and the calibration you specified is

carried out.

To close “About ZERO Calibration”

To close “About ZERO Calibration” setting or to cancel this

mode midway, press the ESC key. A previous screen will

return.

EXAMPLE

Whether upon selection or at once can be determined for each

CH (component).

•Setting upon selection

Select the CH (component) and then perform zero

calibration on the manual zero calibration

screen.

•Setting at once

At a manual zero calibration, CHs (components) for which “at once” was selected can

simultaneously be calibrated.

* When the zirconia O2 analyzer uses the cylinder air or atmospheric air for the zero gas, select “At once”.

Manual Calibration screen