Model 7500 Instruction Manual

Teledyne Analytical Instruments

Terminal block 3 <TN3>

Between 1 – 4: For internal connection. Must not be

wired. (Must not be used as junction

terminal.)

Between 5 – 6: Remote hold input. No hold when open.

Output hold when short-circuited.

Between 7 – 8: Average value reset input. Short-

circuiting the contact input (for at 1.5 sec

or more) resets O2 average and O2

correction average simultaneously.

Opening it restarts the average value.

Between 9 – 10: Automatic calibration remote start

input. Open input after strapping for

at least 1.5 seconds starts the automatic

calibration whether automatic calibration

setting is ON or OFF.

Between 11 – 12: Conductive when analyzer unit error

is producted. Normally open.

Between 13 – 14: CH4 range identification signal

Between 15 – 16: CH3 range identification signal

Between 17– 18: CH2 range identification signal

Between 19– 20: CH1 range identification signal

Note: Range identification signal is conductive at Low range

or open at High range. In case of 1-range system, the signal

remains open.