Model 7500 Instruction Manual

Teledyne Analytical Instruments

8) Description on terminal block

Terminal block 1 <TN1>

Terminal block for analog output (non-isolated output)

Output : 4 to 20 mA or 0 to 100V DC

Between 1 – 2:

CH5 output

Between 3 – 4:

CH4 output

Between 5 – 6:

CH3 output

Between 7 – 8:

CH2 output

Between 9 – 10:

CH1 output

Between 11 – 14:

For internal connection. Must not be

wired. (Must not be used as junction

terminal.)

Between 15 – 16:

CH8 output

Between 17– 18:

CH7 output

Between 19 – 20:

CH6 output

Terminal block 2 <TN2>

Between 1– 2:

For O2 sensor input. (Input for our

Zirconia oxygen sensor or external

O2 sensor. Must not be used unless

O2 meter is added.)

Between 3 –12:

For internal connection. Must not be

wired. (Must not be used as junction

terminal.)

Between 13 – 14:

CH4 remote range changeover input

Between 15 – 16:

CH3 remote range changeover input

Between 17 – 18:

CH2 remote range changeover input

Between 19 – 20:

CH1 remote range changeover input

Note: High range is selected when open. Low range is

selected when short-circuited. For details of action, refer to “6.7

Parameters Setting, “Remote Range”.