31

Model 7500 Instruction Manual

Teledyne Analytical Instruments

6. SETTING AND

CALIBRATION

6.1 Changeover of range

This mode is used to select the ranges of measured

components.

1. During measurement, press the MODE key to display

the User mode.

2. Point the cursor to “Changeover of Range”. Press the

ENT key.

3 The “Channel Selection” screen appears. Press the UP

or DOWN key until the cursor moved selects a desired

CH (component).

4. After selection, press the ENT key.

Note: The range of O2 correction instantaneous values

and O2 correction average values is automatically

switched by changing the range of instantaneous value

of each CH (component).

5. In the Range Setting screen that appears, move the

cursor by pressing the UP or DOWN key to select the

range. (The range with a TRIANGLE mark is currently

selected.)

6. After selection, press the ENT key.

7. Measurement is conducted within the selected range.

The range identification signal (CO) is shorted with the

low range (Range 1), and open with the high range

(Range 2).

Note: If the Remote Range is set to ON, the changeover

of range cannot be performed on the screen.

To close Changeover of range

To close Changeover of range, or cancel the command

midway, press the ESC key. A previous screen will

return.