Installation

11

Figure 2-5. CW-90/CW-90X CPU Board

Table 2-5. Jumper Descriptions

Jumper

Description

JP1/JP2

Jump excitation to sense. If using a 4-wire load cell cable, leave JP1
and JP2 on. If using a 6-wire load cell cable, take JP1 and JP2 off.
Default is ON.

JP3/JP4

Used when upgrading firmware. The jumpers should be on when
upgrading firmware and off when the update is complete.

PC

Power control. If the jumper is set to SW, the POWER key can be used
to turn the unit on/off. If set to ON, the unit will power on when
plugged in and can only be powered off by unplugging.

JMP1

If set to Audit ON, calibration and configuration can be acessed
through the front keypad. If set to Audit OFF, calibration and
configuration can only be accessed by removing the screw on the
base of the unit and pressing the Setup switch with a screwdriver.
Default is Audit ON.

Note

If the RESET button on the CPU board is pressed, the indicator will perform a reboot.