6 annunciators leds – Rice Lake Static Weighing User Manual

Page 8

Advertising
background image

4

CW-90/90X Checkweigher

1.6

Annunciators LEDs

Table 1-2. Annunciator LEDs

LED

Description

0

Zero (Center of Zero) LED

While in gross weight display mode, this LED indicates that the current displayed weight reading is within
+/-0.25 display divisions of the acquired zero, or is within the center of zero band.
When in the net weight display mode, it indicates that the current net weight reading is within +/-0.25 display
divisions of the center of net zero. A display division is the resolution of the displayed weight value, or the
smallest incremental increase or decrease that can be displayed or printed.
G/B LED
Gross weight mode (or Brutto in OIML mode)
N LED
Net weight display mode.
T LED
Indicates that a tare has been acquired and stored by the system.
PT LED
Indicates that a preset tare weight has been keyed in or entered via the EDP serial port.

Displays which unit of measure is being used.
% LED
Percent mode
kg LED
Kilograms
g LED
Grams
lb LED
Pounds
oz LED
Ounces