CW-90/90X Checkweigher

4.0

Calibration

The CW-90/CW-90X can be calibrated using the front panel, EDP commands, or Revolution

®

.

Note

Calibration can be performed in two places within the menu: the CALIBR menu shown in Figure 4-1 and the
SCALE submenu shown in Figure 3-5 on page 18.
The CALIBR menu is a “quick access” calibration; for more
in-depth scale setup and calibration, use the menus found under SETUP

»

CONFIG

»

SCALE.

Calibration consists of the following steps:

Zero calibration

Entering the test weight value

Span calibration

Optional five-point linearization

Optional rezero calibration for test weights
using hooks or chains.

Note

The CW-90/CW-90X

requires the WZERO and WSPAN points to be calibrated. The linearity points are optional,

but must NOT duplicate zero or span. During calibration, the ENTER key acts as a data entry confirmation key.
It also acts as an EXECUTE key, and accepts the value if calibration was successful.

WZERO

Calibrating, please wait

Calibrating, please wait

Display and edi

Press Enter

to calibrate zero

t

test weight value

WVAL

WSPAN

Press Enter to

remove offset from

zero and span

calibrations

REZERO

WLIN

POINT–> 1

POINT–> 2

POINT–> 3

POINT–> 4

POINT–> 5

Same as POINT-> 1

WGT 1

number

Calibrating, please wait

Press Enter

to calibrate span

EXIT

T&D

TEST

SETUP

AUDIT

XXXXXXX

CALIBR

BRIGHT

Press Menu

to cancel

Press Menu

to cancel

Previous A/D raw

counts are shown

Previous A/D raw

counts are shown

ACCUM

Only displayed if

turned on

from Setup menu

CAL1

ID

Only displayed if

turned on

from Setup menu

Figure 4-1. Calibration (CALIBR) Menu

4.1

Front Panel Calibration

1. Press the

MENU

button, then press until the

display reads

CALIBR

(see Figure 4-1).

2. Press

Enter

or to go to the

WZERO

parameter.

3. Press

Enter

or to view the A/D count.

Press

Enter Key to Calibrate WZERO.

Note

Note

Note

If do not want to calibrate zero, press Menu to
exit.

4. The indicator displays

CALIBRATING, PLEASE

WAIT

while calibration is in progress. When

complete, you are advanced to the next prompt
(

WVAL

).

To view the new A/D count, repeat Step 3, but
instead of pressing Enter while viewing the
value, press Menu to exit.

5. With

WVAL

displayed, press

Enter

to display the

stored calibration weight value. Use the
procedure shown in Figure 4-2 to enter the

value of the test weights used for the span
calibration. Press Enter to store the

WVAL

value

and advance to

WSPAN

.

6. With

WSPAN

displayed, press

Enter

or to

view the A/D count. Place test weights on the
scale equal to WVAL. Press Enter to calibrate
WSPAN.

If do not want to calibrate span, press Menu to
exit.

7. After

Enter

is pressed, the indicator displays

CALIBRATING, PLEASE WAIT

. When complete,

you are advanced to the next prompt (

WLIN

).

8. To view the new A/D count, navigate back to

the WSPAN and repeat step 6; however,
instead of pressing Enter while viewing the
value, press Menu to exit.