Calibration

37

0

0 0 0 0 0

When editing numeric values, press or to change the selected digit
Press or to increment or decrement the value of the selected digit
If not already on the screen, the decimal point will flash on the far right.
To move the decimal, press Enter, then use or to move the decimal
position. Press Enter when done.

Figure 4-2. Editing Procedure for Numeric Values

9. Five-point linearization (using the

WLIN

parameter) provides increased scale accuracy
by calibrating the indicator at up to five
additional points between the zero and span
calibrations.

Linearization is optional: if you choose not to
perform linearization, skip the

WLIN

parameter;

if linearization values have previously been
entered, these values are reset to zero during
calibration. To perform linearization, follow
the procedure below.

Note

The linearity points must be less than the
WSPAN point.

With

WLIN

displayed, Press to go to the

first linearization point (

PT-> 1

).

Place test weights on the scale and press

Enter

.

Use the procedure shown in Figure 4-2 to
enter the actual test weight value.The
indicator will display

Press Enter Key to

Calibrate WLIN

.

Press

Enter

to calibrate. The indicator

displays

Calibrating, Please Wait

while

calibration is in progress. When complete,
the A/D count for the linear calibration is
displayed. Press

Enter

again to advance to

the next prompt (

PT-> 2

).

Repeat for up to five linearization points.
To exit the linearization parameters, press

to return to

WLIN

.

10. The rezero function is used to remove a

calibration offset when hooks or chains are
used to hang the test weights. If no other
apparatus was used to hang the test weights
during calibration, remove the test weights and
press to return to the CALIBR menu.

With REZERO displayed, press

Enter

or

to access the rezero function.

If hooks or chains were used during
calibration, remove these and the test
weights from the scale. The indicator will
display

Press Enter Key to REZERO

.

With all weight removed, press

Enter

to

rezero the scale. This function adjusts the
zero and span calibration values. The
indicator displays

Calibrating, Please Wait

while the zero and span calibrations are
adjusted. When complete, the adjusted A/

D c o u n t f o r t h e z e r o c a l i b r a t i o n i s
displayed. Press

Enter

, then press to

return to the CALIBR menu.

11. Press or the

Menu

button to return to weigh

mode.

4.2

EDP Command Calibration

To calibrate the indicator using EDP commands, the
indicator EDP port must be connected to a terminal or
personal computer. See Section 2.7.1 on page 8 for
EDP port pin assignments.

Once the indicator is connected to the sending device,
do the following:

1. Place the indicator in config mode (display

must read

CONFIG - see

Figure 3-5 on page 18)

and remove all weight from the scale platform.
If your test weights require hooks or chains,
place the hooks or chains on the scale for zero
calibration.

2. Send the SC.WZERO command to calibrate

zero. The indicator displays

CALIBRATING,

PLEASE WAIT

while calibration is in progress.

3. Place test weights on the scale and use the

SC.WVAL command to enter the test weight
value in the following format:

SC.WVAL=nnnnnn<CR>

4. Send the SC.WSPAN command to calibrate

span. The indicator displays

CALIBRATING,

PLEASE WAIT

while calibration is in progress.

5. Up to five linearization points can be

c a l i b r a t e d b e t w e e n t h e z e r o a n d s p a n
c a l i b r a t i o n v a l u e s . U s e t h e f o l l o w i n g
commands to set and calibrate a single
linearization point:

SC.WLIN.V1=nnnnn<CR>
SC.WLIN.C1<CR>

The SC.WLIN.V1 command sets the test
weight value (nnnnn) for linearization point 1.
The SC.WLIN.C1 command calibrates the
point. Repeat using the SC.WLIN.Vx and
SC.WLIN.Cx commands as required for
additional linearization points.

6. To remove an offset value, clear all weight

from the scale, including hooks or chains used
t o h a n g t e s t w e i g h t s , t h e n s e n d t h e
S C . R E Z E R O c o m m a n d . T h e i n d i c a t o r
displays

CALIBRATING, PLEASE WAIT

while the

zero and span calibrations are adjusted.

7. Send the SC.KUPARROW or SC.KEXIT EDP

command to return to normal mode.