CW-90/90X Checkweigher

4.3

Revolution

®

Calibration

To calibrate the indicator using Revolution, the
indicator EDP port must be connected to a PC running
the Revolution configuration utility.

1. Place the indicator in config mode (display

reads

CONFIG, see

Figure 3-5 on page 18) and

remove all weight from the scale platform.

2. From Revolution, select

File

»

New

.

The Select Indicator dialog box appears.

3. Select CW-90 and click

OK

.

4. From the Communications menu, select

Connect

.

5. From the left pane, expand the Scale selection

and click the

Scale

button.

Figure 4-3. Scale Button

6. From the Tools menu, select

Calibration Wizard

.

7. Click

NEXT

to begin the Calibration Wizard.

8. Select whether you want to perform a standard

calibration or a standard with multi-point
linearization and click

NEXT

.

9. In the text box, enter the test weight value be

used for span calibration.

10. Select the checkbox if you are using chains or

hooks during the calibration, then click

NEXT.

11. Remove all weight from the scale and click

Click to Calibrate Zero

to begin zero calibration.

If your test weights require hooks or chains,
place them on the scale for zero calibration.

12. When zero calibration is complete, the

Calibration Wizard prompts you to place test
weights on the scale. Place the test weights on
the scale, then click

Click to Calibrate Span

.

13. If you chose to perform linear calibration, the

Calibration Wizard now displays prompts
(1–5). Enter the weight value for Linear Point
#1, place test weights on scale and click

GO

.

Repeat for additional linearization points, then
click

NEXT

.

14. The new and old calibration settings are

displayed. To accept the new values, click

Finish

. To exit and restore the old values, click

Cancel

.

4.4

More About Calibration

The following topics provide additional information about
compensating for environmental factors (Section 4.4.1)
and diagnostic information for determining expected zero
and span coefficients.

4.4.1

Adjusting Final Calibration

Calibration may be affected by environmental factors
including wind, vibration, and angular loading. For
example, if the scale is calibrated with 1000 lb, a strain
test may determine that at 2000 lb the calibration is 3 lb
high. In this case, final calibration can be adjusted by
tweaking WVAL to 998.5 lb. This adjustment provides
a linear correction of 1.5 lb per 1000 lb.

To adjust the final calibration, return to the

WVAL

prompt and press

ENTER

to show the test weight value.

Press or to adjust calibration up or down. Press

ENTER

to save the value, then press to return to the

CALIBR menu.