25

Level 3 submenus

MODE

RANGE
TARGWT
TARG%

Sets checkweighing mode to range, target weight, or target percent. See Section 5.0 on
page 39
for more information.

VALSET

KEYED
PUSH
BOTH
RDONLY

KEYED allows the operator to digitally set the Accept band tolerance values using keypad
buttons while in Normal Weighing mode.

PUSH prohibits the operator from digitally entering values with the keypad. Requires the
operator to place actual weights on the scale, then press keypad buttons to acquire lower and
upper Accept band tolerance values.

BOTH is a combination mode that allows the operator to have the CW-90/CW-90X acquire
Accept band tolerances from actual weights on the scale, but then gives the operator an
opportunity to digitally modify those values directly afterward.

RDONLY is a “read only” mode which allows the operator to see but not change values.

NEGCKW

OFF
ON
AUTO

Sets whether negative checkweighing mode is off, on, or is using auto-tare.

ID

OFF
ON

Turns IDs on or off. See Section 5.4 on page 43 for more information on IDs. if IDs are turned
off, the ID selection will not appear in the main menu structure.

DATA

TYPE
PORT

Selects the data type and port number for that data type.

LIGHTS

SOLID
GRAPH
OFF

Select SOLID if you want the checkweigher to display a solid light bar for over and under
weights; select GRAPH if you want it to illuminate lights in increments relative to how close the
weight is the acceptance band; Select OFF if you want the lights to never be illuminated.

See Section 1.7 on page 5 for an illustration of graph LEDs.

REGULA

NTEP
CANADA
REGULA
NONE
OIML

Regulatory mode. Specifies the regulatory agency having jurisdiction over the scale site.

Note

The value specified for REGULA affects the function of the front panel TARE
and ZERO keys.

OIML, NTEP, and CANADA modes allow a tare to be acquired at any weight greater
than zero. NONE allows tares to be acquired at any weight value.

OIML, NTEP, and CANADA modes allow a tare to be cleared only if the gross weight
is at no load. NONE allows tares to be cleared at any weight value.

NTEP and OIML modes allow a new tare to be acquired even if a tare is already
present. In CANADA mode, the previous tare must be cleared before a new tare can
be acquired.

NONE, NTEP and CANADA modes allow the scale to be zeroed in either gross or net
mode as long as the current weight is within the specified ZRANGE. In OIML mode,
the scale must be in gross mode before it can be zeroed; pressing the ZERO key in
net mode clears the tare.

INDUST provides a set of subparameters to allow customization of tare, clear, and
print functions in non-legal-for-trade scale installations.

REGWOR

GROSS
BRUTTO

Sets the term displayed when weighing in gross mode. Selecting BRUTTO replaces the Gross
annunciator with Brutto.

DECFMT

DOT
COMMA

Specifies whether decimal numbers are displayed using a period (DOT) or a comma.

TIME

TFORMT
TSEP
SET

Allows you to set the current time, and the time format and separator character.

DATE

DFORMT
D SEP
SET

Allows you to set the current date, and date format and date separator character.

FEATUR Menu

Parameter

Choices

Description