62

CW-90/90X Checkweigher

9.4

Continuous Output (Stream) Format

Figure 9-1 shows the continuous output format sent to a CW-90/CW-90X port when that port’s TRIGGE parameter
(SERIAL menu) is set to STREAM.

<STX>

<UNIT>

<G/N>

<S>

<TERM>

ASCII 02
(decimal)

Weight data: 7 digits, right-justified, with
decimal point, leading zero supression.

Overload = &&&&&&&
Underrange = = = = = = = =
Display overflow = OVERFL

L = pounds
K = kilogram
G = grams
O = ounces

G = Gross
N = Net

Status:
<space > = valid
I = Invalid
M = Motion
O = Over/under range

Polarity:

< space > = Positive

<–> = Negative

<^> = Overload

<]> = Underrange

<CR> <LF>

or <CR>

ASCII 13, 10

(decimal)

<space> = lb/oz

<wwwwwww>

<POL>

Figure 9-1. Continuous Output Data Format

9.5

Digital Filtering

The CW-90/CW-90X uses averaged digital filtering to reduce the effect of vibration on weight readings. Adjustable
threshold and sensitivity functions allow quick settling by suspending filter averaging, allowing the weight reading
to jump to the new value. Figure 9-2 shows the digital filter parameters on the CONFIG menu.

DFLTR1

64

128

4

16

8

32

256

DFSENS

DFTHRH

32OUT

64OUT

2OUT

8OUT

4OUT

16OUT

128OUT

20D

50D

250D

5D

2D

10D

100D

NONE

200D

1

2

DFLTR2

64

128

4

16

8

32

256

1

2

DFLTR3

64

128

4

16

8

32

256

1

2

Figure 9-2. Digital Filtering Parameters on the Configuration (CONFIG) Menu

DIGFL1= n1

DIGFL2= n2

DIGFL3= n3

3rd Stage

2nd Stage

1st Stage

Filter Average

Displayed

Value

n2

1st Stage

Filter Averages

n1

A/D Readings

n3

2nd Stage

Filter Averages

Filter Average

Filter Average

Figure 9-3. Flow Diagram for CW-90/CW-90X Digital Filters