12 SCT Weight Transmitter Operator’s Manual

2.1

Calibration

Figure 2-2. Calibration Menu Structure

Parameter

Choices

Description

FS-tEO

Enter #

0 =dENO *

System Full Scale is determined by multiplying one load cell capacity by the number of load cells
used. Example of system full scale value calculation:

4 cells of 1000kg ---->

FULL SCALE = 1000 X 4 = 4000

The instrument is supplied with a theoretical full scale value deno corresponding
to 10000. To restore factory values, set 0 as full scale.

SEnSI b

Enter #

0.50000

to

7.00000

2.00000 *

Sensitivity is a load cell rated parameter expressed in mV/V. Set the average sensitivity value
indicated on the load cells. Example of 4-cell system with sensitivity

2.00100, 2.00150, 2.00200, 2.00250;

enter 2.00175,

calculated as (2.00100 + 2.00150 + 2.00200 + 2.00250) / 4.

dI UI S

1
2

*

5
10
20
50
100
0.0001
0.0002
0.0005
0.001
0.002
0.005
0.01
0.02
0.05
0.1
0.2
0.5

Division (resolution) - the weight increment (display division size) that the scale counts by.
Selections are: 0.0001 and 100 with x1 x2 x5 x10 increments.

dI UI S

WEIGHt

unIt

COEFF

LItrE

bAr

AtN

PI ECE

nEUton

lb

OtHEr

HI LOG

HI LO-N

nEU-N

G
t

FS-tEO

Enter #

SEnS Lb

Enter #

NASS

Enter #

ZErO

Enter #

Enter #

Enter #

I NP 0

Enter #

50

100

0.0001

0.0002

20

10

0.002
0.005

0.001

0.0005

2
5

0.02
0.05

0.01

0.2
0.5

0.1

1

FILtEr

PArA 0

SErIAL

tESt

Out-In

ANA LOG

000000

SEtP1

CALIb

PtArE

Note