Configuration

19

2.5

Analog Output

Figure 2-3. Analog Menu Structure

Soldered Jumper

For the output -10 +10 V and -5 +5 V the soldered jumper
(SW4) must be closed:

Release the locking tabs, both sides, to open
enclosure.

Locate the soldered jumper (SW4), on the circuit
board. See Figure 2-4.

Close the jumper shorting the pads with a solder
bridge.

Parameter

Choices

Description



4-20 mA

*

0-20 mA
0-10 V
0-5 V
-10 +10 V
-5 +5 V

Selects the analog output type.

See “Soldered Jumper” on page 19
See “Soldered Jumper” on page 19



Enter #




Select mode to be tracked, gross or net. If the net function is not active, the analog output varies
according to gross weight.



Enter #

Set the weight value for the minimum analog output value.

Only set a value different from zero to limit the analog output range.

For instance: for a full scale value of 10000 kg, a 4 mA signal at 5000 kg is required,
and 20 mA at 10000 kg, in this case, instead of zero, set 5000 kg.



Enter #

Set the weight value for the maximum analog output value; it must correspond to the value set in the
PLC program (default: calibration full scale).

E.g.: if using a 4-20 mA output and in the PLC program a 20 mA = 8000 kg is desired, set the
parameter to 8000.



Analog output correction to zero: if necessary adjust the analog output, allowing the PLC to indicate 0.
The sign ‘-‘ can be set for the last digit on the left.

E.g.: For a 4-20 mA output and a minimum analog setting, the PLC or tester reads 4.1 mA. Set
the parameter to 3.9 to obtain 4.0 on the PLC or tester. (See “Analog Output Type Scale
Corrections” on page 20)



Full scale analog output correction: if necessary adjust the analog output, allowing the PLC to indicate
the value set in the parameter.

E.g. For a 4-20 mA output with the analog set to full scale and the PLC or tester reads 19.9 mA,
set the parameter to 20.1 to obtain 20.0 on the PLC or tester. (See “Analog Output Type Scale
Corrections” on page 20)

* Indicates default value.

),/W(U

3$U$

6(U,$/

W(6W

2XW,Q

$1$/2*



3W$U(

6(W3

&$/,E

P2G(

W<3(

Enter #

X

P$

P$

$Q$

$Q$)6

&2U)6

&2U

Enter #

Enter #

Enter #

Enter #

X



X

*URVV

1HW

Note

Figure 2-4. Soldered Jumper