Analog output setting (analog models only) – Rice Lake Signal Conditioning Transmitter/Indicators User Manual

Page 53

Advertising
background image

Operation 49

At the time of writing, the setpoints, hysteresis values are saved to the RAM and will be lost upon the next
power-off; to store them permanently to the EEPROM so that they are maintained at power-on, the ‘99’
command of the Command Register must be sent.

Weight (Span) Calibration Commands (With Test Weights)

The instrument calibration can be changed via MODBUS. To carry out this procedure, the system must be

unloaded and the weight value display reset to zero with the command ‘100’ of the Command Register. Then, a

load must be placed on the system and the correct weight value must be sent to the registers 40037-40038; to save

this value, send the control ‘101’ from the Command Register. If the operation is successfully completed, the two

test weight registers are set to zero.

Analog Output Setting (Analog Models Only)

Write the weight in the registers “Weight value corresponding to the Full Scale of analog output H” (40045) and

“Weight value corresponding to the Full Scale of analog output L” (40046) or write the weight in the registers

“weight value corresponding to the ZERO of the analog output H” (40043) and “weight value corresponding to the

ZERO of the analog output L” (40044). After writing the value, send the command 99 from the Command Register

to save it to EEPROM memory.

40017

SETPOINT 1 H

Only after command ‘99’ of the

COMMAND REGISTER

R/W

40018

SETPOINT 1 L

R/W

40019

SETPOINT 2 H

R/W

40020

SETPOINT 2 L

R/W

40021

SETPOINT 3 H

R/W

40022

SETPOINT 3 L

R/W

40023

HYSTERESIS 1 H

R/W

40024

HYSTERESIS 1 L

R/W

40025

HYSTERESIS 2 H

R/W

40026

HYSTERESIS 2 L

R/W

40027

HYSTERESIS 3 H

R/W

40028

HYSTERESIS 3 L

R/W

40029

INPUTS

-

R

40030

OUTPUTS

NO

R/W

40037

Test weight for calibration H

Use with command ‘101’ of the

COMMAND REGISTER

R/W

40038

Test weight for calibration L

R/W

40043

Weight value corresponding to ZERO of the analog
output H

Only after command ‘99’ of the

Command Register.

(Analog Models Only)

R/W

40044

Weight value corresponding to ZERO of the analog
output L

R/W

40045

Weight value corresponding to Full Scale of the analog
output H

R/W

40046

Weight value corresponding to Full Scale of the analog
output L

R/W

REGISTER

DESCRIPTION

Saving to EEPROM

ACCESS

Important

Advertising