48 SCT Weight Transmitter Operator’s Manual

Communication Error Management

The communication strings are controlled by CRC (Cyclical Redundancy Check).
In case of a communication error the slave will not respond with any string. The master must allow for a time-out

before response reception. If no response is received it infers that a communication error has occurred.
In the event of a string received correctly but not executable, the slave responds with an EXCEPTIONAL
RESPONSE. The "FUNCTION" field is transmitted with the msb at 1.

List Of Usable Registers

The MODBUS-RTU protocol implemented on this instrument can manage a maximum of 32 registers read and

written in a single query or response.

R

= the register can be read only

W

= the register can be written only

R/W = the register can be both read and written
H

= high half of the DOUBLE WORD forming the number

L

= low half of the DOUBLE WORD forming the number

EXCEPTIONAL RESPONSE

Address

Function

Code

2 bytes

A

Funct + 0x80

CRC

CODE

DESCRIPTION

1

ILLEGAL FUNCTION (Function not valid or not supported)

2

ILLEGAL DATA ADDRESS (The specified data address is not available)

3

ILLEGAL DATA VALUE (The data received have no valid value)

REGISTER

DESCRIPTION

Saving to EEPROM

ACCESS

40001

Firmware version

-

R

40002

Type of instrument

-

R

40003

Year of Production

-

R

40004

Serial Number

-

R

40005

Active program

-

R

40006

COMMAND REGISTER

NO

W

40007

STATUS REGISTER

-

R

40008

GROSS WEIGHT H

-

R

40009

GROSS WEIGHT L

-

R

40010

NET WEIGHT H

-

R

40011

NET WEIGHT L

-

R

40012

PEAK WEIGHT H

-

R

40013

PEAK WEIGHT L

-

R

40014

Divisions and Units of measure

-

R

40015

Coefficient H

R

40016

Coefficient L

R