32 SCT Weight Transmitter Operator’s Manual

2.13.5Program Selection - Reverse:

Scale capacity is displayed when scale is empty. As weight is added display will count down.

1. With power off, press and hold

, then power on. PrOG is displayed.

2. Press ,

bASE is displayed.

3. Press ,

rEuEr is displayed.

4. Press ,

UAIt is displayed.

5. Instrument will reboot.

2.13.6Program Selection - Not Legal:

Scale approval state is preset for not-legal. For other choices please contact technical assistance:

nOtLEG: Not Legal for Trade.
LEGAL: Legal for Trade (OIML)
NuLt: Legal for Trade, multi-interval (OIML)

1. With power off, press and hold

, then power on. PrOG is displayed.

2. Press ,

bASE is displayed.

3. Press ,

NOtLEG is displayed.

4. Press ,

UAIt is displayed.

5. Instrument will reboot.

By pressing

you will quit the program without introducing any changes and without deleting any of the set

variables.

If you do not have a specific manual for the newly set program, you can request it from technical assistance.

2.13.7Keypad Or Display Locking

1. Press

immediately followed by

, hold them down for about 5 seconds (this operation is also possible

via the MODBUS and ASCII protocols):

2. Press

or

until desired parameter is displayed, press

.

Parameter

Description

FrEE

no lock

KEY

keypad lock: if active, when key is pressed the message bLOC is
displayed.

dI SP

Keypad and Display lock: if active, the keypad is locked and the
display shows the instrument model (weight is not displayed); by
pressing a key the display shows bLOC for 3 seconds.

Note