3 load cells, 1 load cell input test (quick access), 2 load cell testing – Rice Lake Signal Conditioning Transmitter/Indicators User Manual

Page 7

Advertising
background image

Introduction

3

Windy Conditions - Shocks - Vibrations:

The use of weigh modules is strongly recommended for all load cells to compensate for misalignment of the

support surfaces. The system designer must ensure that the scale is protected against lateral shifting and tipping

relating to shocks and vibration, windy conditions, seismic conditions and stability of the support structure.

Grounding The Weighed Structure:

By means of a 10ga solid or braided wire or braided grounding strap, connect the load cell upper support plate with

the lower support plate, then connect all the lower plates to a single earth ground. Once installed electrostatic

charges accumulated are discharged to the ground without going through or damaging the load cells. Failure to

implement a proper grounding system might not affect the operation of the weighing system; this, however, does

not rule out the possibility that the load cells and connected instrument may become damaged by ESD. It is

forbidden to ensure grounding system continuity by using metal parts contained in the weighed structure.(see

Figure 1-1.)

Figure 1-1. Installation Recommendations

1.3

Load Cells

1.3.1 Load Cell Input Test (Quick Access)

1. From the weight display, press

for three seconds.

2. The display will read NU-CEL. Press

.

3. The response signal of the load cell is displayed, expressed in mV with four decimals. Press

three times

to exit test mode.

1.3.2 Load Cell Testing

Load Cell Resistance Measurement (Use A Digital Multimeter):

• Disconnect the load cells from the instrument and check that there is no moisture in the load cell junction box

caused by condensation or water infiltration. If so, drain the system or replace it if necessary.

• The value between the positive signal wire and the negative signal wire must be equal or similar to the one

indicated in the load cell data sheet (output resistance).

• The value between the positive excitation wire and the negative excitation wire must be equal or similar to the

one indicated in the load cell data sheet (input resistance).

• The insulation value between the shield and any other load cell wire and between any other load cell wire and

the body of the load cell must be higher than 20 Mohm (mega ohms).

Uses ground plate to

continue ground.

Uses structure to continue ground.