Rice Lake Signal Conditioning Transmitter/Indicators User Manual

Page 27

Advertising
background image

Configuration 23

Front Panel Configuration

1. Press and hold

and

to enter set-up menu. CALIb will be displayed.

2. Press

or

until EtHnEt is displayed, press

.

3. Press

or

until desired parameter is displayed, press

.

4. Press

or

until desired value is displayed, press

.

5. Press

twice to exit set-up menu.

NodE

(Continued)

Hdr1 P

Continuous weight transmission protocol, streams net and gross including decimal. (see
Section 3.14) (set: BAUD=9600,PARITY=none, STOP=1)

Hdr1 PN

Continuous weight transmission protocol (see Section 3.14)
(set: BAUD=9600,PARITY=none, STOP=1)

When the remote display is set to gross weight:
- if the instrument displays the gross weight, the remote display shows the gross weight.
- if the instrument shows the net weight the remote display shows the net weight alternated
with the message “net”

bAud

2400
4800
9600 *

1

9200

38400
115200

Transmission speed.

The baud rate must be the same as the baud rate setting in the Serial
Communications Settings, see Section 2.9.

ADDR

1-99
1 *

Instruments address

HERTZ

Maximum Transmission Frequency
To be set when the CONTIN transmission protocol is selected. (see Figure 2-7)

1

0Hz *

20Hz

Max setting with min 2400 baud rate

30Hz

40Hz

Max setting with min 4800 baud rate

50Hz

60Hz

70Hz

80Hz

Max setting with min 9600 baud rate

100Hz

Max setting with min 19200 baud rate

200Hz

Max setting with min 38400 baud rate

300Hz

Max setting with min 38400 baud rate

dELAY

0-200 msec

0 *

Delay in milliseconds which elapses before the instrument replies

PAritY

nOnE *
EUEn
Odd

parity none
even parity
odd parity

StOP

1 *

2

Stop bit

* - indicates default value.
** - These settings do not function with Firmware 1.02.00 or lower.

Parameter

Choices

Description

Note

Advertising