SCT Weight Transmitter Operator’s Manual

Load Cell Voltage Measurement (Use A Digital Multimeter):

• Remove weight of scale from load cell to be tested.
• Make sure that the excitation wires of the load cell connected to the instrument is 5 Vdc +/- 3%.
• Measure the millivolt signal between the positive and the negative signal wires by directly connecting them to

the multi-meter, and make sure it reads between 0 and 0.5 mV (thousandths of a Volt).

• Apply load to the load cell and make sure that there is a signal increment.

If one of the above conditions is not met, please contact the technical assistance service.

1.4

Specifications

• Weight indicator and transmitter for Omega/DIN rail mounting suitable for back panel; space saving vertical

shape. Six-digit semi alphanumeric display (18mm h), 7 segment. Four-key keyboard. Dimensions:

25x115x120 mm.

• Displays the gross weight; with an external contact capable of remote zeroing and gross/net switching.
• Peak weight function.
• Analog Output Option - Transmits the gross or net weight via opto-isolated analog output 16 bit, current 0-

20mA, 4-20mA or voltage 0-10V, 0-5V (±0V / ±5V by closing a soldered in jumper).

DeviceNet Option – It has a DeviceNet DP V0 Slave port that allows to exchange the main system parameters

with a DeviceNet master.
Profibus Option – It has a Profibus DP V0 Slave port that allows to exchange the main system parameters

with a Profibus master.
Ethernet Option – It has an Ethernet/IP device port that allows to exchange the main system parameters with

an Ethernet/IP scanner.
Modbus/TCP Option – It has a Modbus/TCP device port that allows to exchange the main system parameters

with a Modbus/TCP master.

• Transmits the gross or net weight via RS-485 serial port, by means of protocols:

- Modbus RTU
- ASCII bidirectional protocol
- Continuous transmission

Important

ANALOG OUTPUT

(Current and Voltage)

PLC

PLC or FIELD SIGNALS

2 DIGITAL INPUTS

(Optoisolated,

Externally supplied)

JUNCTION BOX

3 RELAY

MAX 8 LOAD CELLS IN

PARALLEL

RS-485
RS-232
Modbus RTU

SERIAL PORT

DC power

supplier

(12-24 Volt )

Other Options

Devicenet

Profibus

Ethernet

OUTPUTS