16 SCT Weight Transmitter Operator’s Manual

2.1.6 Setting Units of Measure

1. Press and hold

, then press

to enter set-up menu, CALIb will be displayed.

2. Press ,

FS-tEO is displayed.

3. Press

or

until unIt is displayed, press

.

4. Press

or

until desired unit is displayed, press

.

5. Press

twice to exit set-up menu.

* Indicates it is possible to set the display coefficient. To use COEFF it is necessary to enable it, closing the COEFF

input. See Section 2.1.7 on page 16.

If the print function is enabled, the symbol of the selected unit of measure will be printed after the measured
value.

2.1.7 Display Coefficient

By setting the coefficient the display is changed accordingly.
If one of the inputs is set to COEFF mode (see Section 2.10 on page 27) when the input is closed the value will be

displayed modified according to the coefficient; when the input is opened the standard weight display will be

restored.

1. Press and hold

, then press

to enter set-up menu, CALIb will be displayed.

2. Press ,

FS-tEO is displayed.

3. Press

or

until COEFF is displayed, press

.

4. Press

or

until desired number is displayed, press

.

5. Press

twice to exit set-up menu.

All other settings (setpoints, hysteresis, calibration ...) are expressed in weight value. If you want to
convert them to the new unit of measurement, perform one of the following procedures for changing the
system calibration.
The parameter must remain set to 1.0000.

HI LOG

kilograms

LI tre

litres*

G

grams

bAr

bar*

t

tons

AtN

atmospheres*

Lb

pounds*

PI ECE

pieces*

nEUton

newton*

nEU-N

newton metres*

LI tre

litres*

HI LO-N

kikgram metres*

bAr

bar*

OtHEr

units of measure not included*

HI LOG

kilograms

G

grams

t

tons

Lb

pounds

Value set in COEFF will be multiplied by the weight value currently displayed

nEUton

newton

Value set in COEFF will be multiplied by the weight value currently displayed

LI tre

litres

in COEFF set the specific weight in kg/l, assuming that the system is
calibrated in kg

bAr

bar

Value set in COEFF will be multiplied by the weight value currently displayed

AtN

atmospheres

Value set in COEFF will be multiplied by the weight value currently displayed

PI ECE

pieces

in COEFF set the weight of one piece

nEU-N

newton metres

Value set in COEFF will be multiplied by the weight value currently displayed

HI LO-N

kikgram metres

Value set in COEFF will be multiplied by the weight value currently displayed

OtHEr

other generic units of
measure not included in list

Value set in COEFF will be multiplied by the weight value currently displayed

Note

Important