17. Temperature safety devices

17.1 Overtemperature protective device (class 1)

The refrigerated incubator is equipped with an internal temperature safety device, class 1 acc. to DIN

12880. It serves to protect the unit and prevents dangerous conditions caused by major defects.
If a temperature of approx. 110 °C / 230 °F is reached, the over temperature protective device

permanently turns off the unit. The user cannot restart the device again. The protective cut-off device is

located internally. Only a service specialist can replace it. Therefore, please contact an authorized service

provider or BINDER Service.

17.2 Overtemperature safety controller (temperature safety device class 3.1)

The chamber is regularly equipped with an electronic overtemperature safety controller (temperature

safety device class 3.1 according to DIN 12880).
The overtemperature safety controller serves to protect the refrigerated incubator, its environment and the

contents from exceeding the maximum permissible temperature. In the case of an error, it limits the

temperature inside the chamber to the entered safety controller set-point.
Please observe the guideline BGI/GUV-I 850-0 on safe working in laboratories (formerly BGR/GUV-R 120

or ZH 1/119 laboratory guidelines issued by the employers’ liability insurance association) (for Germany).
Set the safety controller set point by approx. 2 °C up to 5 °C above the desired temperature set point.

Recommended setting: Set-point type “Offset” with safety controller set-point 2 °C (factory setting).
The safety controller is functionally and electrically independent of the temperature control system. If an

error occurs, it performs a regulatory function.

Check the setting regularly and adjust it following changes of the set-point or charge.

17.2.1 Safety controller modes

You can select between “Limit (absolute)” and “Offset (relative)” safety controller mode

Limit: Absolute maximum permitted temperature value
Example:
Temperature set point 40 °C
Safety controller set-point 45 °C
This setting offers high safety as a defined temperature limit will not be exceeded. It is important to

adapt the safety controller set-point after each modification of the temperature set-point. Otherwise,

the limit could be too high to ensure efficient protection, or, in the opposite case, it could prevent the

controller from reaching an entered set-point outside the limit range.

Offset: Maximum overtemperature above any active temperature set point. The maximum

temperature changes internally and automatically with every set-point change.
Example:
Temperature set point 40 °C
Safety controller set-point 5 °C
This setting is recommended for program operation. It is important to check the safety controller set-

point and safety controller mode occasionally, as it does not offer a fix, independent limit temperature

value, which would never be exceeded.