KT (E6) 09/2013

Page 113/135

19.6 Analog output for temperature (option)

With this option the chamber is equipped with an analog output 4-20 mA for temperature. This output

allows transmitting data to external data registration systems or devices.
The connection is realized as a DIN socket (3) on the unit rear. A suitable DIN plug is enclosed.

ANALOG OUTPUT 4-20 mA DC
PIN 1: Temperature –
PIN 2: Temperature +
Temperature range: -10 °C / 32°F to +100 °C / 212°F

Figure 10: Pin configuration of the DIN socket (3) on the unit rear

19.7 Zero-voltage relay control outputs (option)

The zero-voltage relay control outputs 1, 2 und 3 are used to switch any device connected via a DIN

socket on the unit rear. They permit turning on and off individually the connected devices by the

controller. They can be programmed in fixed value entry mode (chap. 7.3) as well as in the time program

editor (chap. 9.3.7) or the week program editor (chap. 10.3.6) via the operation lines.
The connection is realized as a DIN socket (4) on the unit rear. A suitable DIN plug is enclosed.

Figure 11: Pin configuration of the DIN socket (4) on the unit rear

Relay output 1

Relay output 2

Relay output 3

Pin 1: Pin

Pin 2: Make

Pin 3: Pin

Pin 4: Make

Pin 5: Pin

Pin 6: Make

Switching state On:

1xx

Switching state On: x

1x

Switching state On: xx

1

Maximum loading capacity of the switching contacts: 24V AC/DC – 2.5 A

DANGER

Electrical hazard.
Danger of death.
Damage to switching contacts and connection socket.

Do NOT exceed the maximum switching load of 24V AC/DC – 2.5A.

Do NOT connect any devices with a higher loading capacity.