Installation of the equipment, 1 spacer for wall distance, Danger – BINDER KT 53 User Manual

Page 21

Advertising
background image

KT (E6) 09/2013

Page 21/135


To completely separate the unit from the power supply, you must disconnect the power plug. Install the

unit in a way that the power plug is easily accessible and can be easily pulled in case of danger.
With an increased amount of dust in the ambient air, clean the Peltier fan grid (7) by suction or blowing

several times a year.
Avoid any conductive dust in the ambiance according to the unit layout complying with pollution degree 2

(IEC 61010-1).
Do not install or operate the unit in potentially explosive areas.

DANGER

Explosion hazard.
Danger of death.

Do NOT operate the unit in potentially explosive areas.

KEEP explosive dust or air-solvent mixtures AWAY from the vicinity of the unit.


4.

Installation of the equipment

4.1 Spacer for wall distance

Please fix both spacers with the supplied screws at the unit rear. This serves to ensure the prescribed

minimum distance to the rear wall of 100 mm / 3.94 in.

Figure 6: Spacer for wall distance

Figure 7: Rear KT with mounted spacers

Advertising
This manual is related to the following products: