BINDER KT 53 User Manual

Page 105

Advertising
background image

KT (E6) 09/2013

Page 105/135


Overview of the current settings
You can check the current settings of the safety controller:

Fixed value

08.03.2013 05:05:06

Submenu “Safety controller”

(view with standard unit equipment)
Select “Show settings”

and press the operating button.

..\ Safety controller
Mode

Setpoints

Show settings

Close

Home

The overview display shows the set-points and actual values of the main temperature controller and the

safety controller and indicates the safety controller mode.

Fixed value

08.03.2013 05:05:06

Overview display with safety controller mode “Limit”

(example values).
In case of a temperature safety device class 3.1

combined with optional unit equipment, “Safety

controller” is indicated as “Overtemperature”

instead.

Setpoint Actual

Chamber temperature [°C]

37.0 37.6

Safety controller mode

Limit (absolute)

Safety controller [°C]

40.0 37.6

Close

Home

Fixed value

08.03.2013 05:05:06

Overview display with safety controller mode

“Offset” (example values).
In case of a temperature safety device class 3.1

combined with optional unit equipment, “Safety

controller” is indicated as “Overtemperature”

instead.

Setpoint Actual

Chamber temperature [°C]

37.0 37.6

Safety controller mode

Offset (relative)

Safety controller [°C]

3.0 37.6

Close

Home

Go back to the initial view with “Home”.

Advertising
This manual is related to the following products: