BINDER KT 53 User Manual

Page 109

Advertising
background image

KT (E6) 09/2013

Page 109/135


Overview of the current settings
You can check the current settings of the safety controller:

Fixed value

08.03.2013 05:05:06

Submenu “Safety controller”
Select “Show safety controller settings”

and press the operating button.

..\ Safety controller
Mode

Over temperature

Under temperature

Show settings

Close

Home


The overview display shows the set-points and actual values of the main temperature controller and the

safety controller and indicates the safety controller mode.

Fixed value

08.03.2013 05:05:06

Overview display with safety controller mode “Limit”

(example values)

Setpoint Actual

Chamber temperature [°C]

37.0 37.6

Safety controller mode

Limit (absolute)

Overtemperature [°C]

40.0 37.6

Undertemperature [°C]

32.0

Close

Home

Fixed value

08.03.2013 05:05:06

Overview display with safety controller mode

“Offset” (example values)

Setpoint Actual

Chamber temperature [°C]

37.0 37.6

Safety controller mode

Offset (relative)

Overtemperature [°C]

3.0 37.6

Undertemperature [°C]

5.0

Close

Home

Go back to the initial view with “Home”.