14.3 Confirming an active alarm

Fixed value

08.03.2013 15:05:02

Initial view with overtemperature safety controller

alarm.
The buzzer sounds (if not deactivated previously).
Press the “Info” button.

Setpoint

Actual value

Temperature [°C]

25.0 29.8

Fan

[%]

100.0

User

Info

Menu

Alarm acknowledgement

08.03.2013 05:05:06

Submenu “Alarms” for alarm acknowledgement.
The active alarm that has been triggered first is

highlighted.
Press the “Reset” button to confirm the alarm and

mute the buzzer.

..\ Home/Alarms

Safety controller overtemperature

Close

Reset

Home

The initial view is displayed. The alarm is still active, therefore the alarm icon continues to be shown. The

buzzer is off.
If there is another active alarm, you can press the “Info” button again and also confirm the alarm with the

“Reset” button.


The following alarm messages may appear in the “Alarm list”:

Alarm message

Signification

Corresponding

alarm icon

Safety controller

overtemperature

Overtemperature safety controller alarm (class 3.1): selected

value of the safety controller exceeded.

Safety controller

overtemperature

Overtemperature safety controller alarm with option class 3.3:

selected value of the safety controller exceeded

Safety controller

undertemperature

Undertemperature safety controller alarm with option class 3.3:

fallen below selected value of the safety controller

Temp. range

Having reached the set-point, the temperature deviates by

more than +/- 2 °C from the entered set-point