BINDER KT 53 User Manual

Page 123

Advertising
background image

KT (E6) 09/2013

Page 123/135


Fault description

Possible cause

Required measures

Heating (continued)

With option safety

controller class 3.3:

undertemperature safety

controller responds.

Alarm messages:

Safety controller set-point value

exceeded.

Check setting of temperature set-

point and of the safety controller

class 3.2 set-point. If appropriate,

select suitable value.

Controller defective.

Contact BINDER service.

Safety controller defective.

Refrigerating performance

Low or no refrigerating

performance.

Notification:

Ambient temperature above

25 °C / 77°F (chap. 3.4).

Select cooler place of installation.

Refrigerating system not turned

on.

Contact BINDER service.

Too much external heat load.

Reduce heat load.

Low or no refrigerating

performance.

Alarm message:

Error in the refrigerating

system.

Contact BINDER service.

Controller

Program duration longer than

programmed.

Inappropriate tolerances have

been programmed.

For rapid transition phases, do

NOT program tolerance limits in

order to allow maximum heating

speed.

Program stops one section too

early.

Program line is incomplete.

When programming with setting

“ramp”., define the end value of the

desired cycle by adding an

additional section with a section

time of at least one minute.

The controller returns to the initial

view from any level.

No button was pressed for

more than 120 sec.

Repeat entries, enter the values

rapidly.

Message “Sensor rupture”

Sensor rupture between sensor

and controller

Contact BINDER service.

Only qualified service personnel authorized by BINDER must perform repair.

Repaired units must comply with the BINDER quality standards.