4.2 Electrical connection

The refrigerated incubator comes with an IEC connector plug.

Model

Art. no.

(x = 0 or 1)

Power plug

Voltage

+/-10 %

Current

type

Power

frequency Unit fuse

KT 53

KT 115

9x20-0250

9x20-0209

Shock-proof

plug

200 V to 240 V

1N~

50/60 Hz

10 A

KT 53-UL

KT 115-UL

9x20-0255

9x20-0227

NEMA 5-15P

100 V to 120 V

1N~

50/60 Hz

10 A

Prior to connection and start-up, check the power supply voltage. Compare the values to the specified

data located on the unit’s type plate (left unit side, bottom right-hand, see chap. 1.4).

When connecting, please observe the regulations specified by the local electricity supply company as

well as the VDE directives (for Germany).

Pollution degree (acc. to IEC 61010-1): 2

Over-voltage category (acc. to IEC 61010-1): II

CAUTION

Danger of incorrect power supply voltage.
Damage to the equipment.

Check the power supply voltage before connection and start-up.

Compare the power supply voltage with the data indicated on the type plate.

See also electrical data (chap. 23.4).

To completely separate the unit from the power supply, you must disconnect the power plug.

Install the unit in a way that the power plug is easily accessible and can be easily pulled in

case of danger.